DUT Output Timing/Jitter - Procedures
This section shows how to set up and use the DUT Output Timing/Jitter
measurement. As an example, we measure the output timing and jitter behavior
of a shielded cable.
This requires the following steps:
• Preparing the measurement (see
“How to Prepare the DUT Output Timing/
Jitter Measurement” on page 250
)
• Executing the measurement (see
“How to Execute the DUT Output Timing/
Jitter Measurement” on page 251
)
• Optimizing the view of the results (see
“How to Optimize the View of the
)
How to Prepare the DUT Output Timing/Jitter Measurement
To use the DUT Output Timing/Jitter measurement to test a shielded cable:
1
Disable the pattern generator outputs by pressing the
0V (Disable)
button in
the
PG Setup -> Data Output
screen.
2
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
3
Terminate all non-connected pattern generator output ports with 50 Ω.
4
Use a shielded cable to connect the pattern generator's Data Out port and the
error detector's Data In port.
5
Switch to the
Pattern
menu and press
Pattern Select
. Select an appropriate
pattern, for example a pure 2^15-1 PRBS, for both the pattern generator and
the error detector.
6
For the pattern generator setup you need to specify the logic levels and the bit
rate. Select
LVPECL
levels and
Gb Ethernet
(1.06250Gb/s) as clock speed. See
“Setting up the Pattern Generator - Concepts ” on page 121
for more
information.
7
Set up the error detector so that the input range and the termination matches
the pattern generator's levels:
– Select an
Input Range
from 1V to 3V
– Set the
Data Termination
to 1.3V
– Set the
Clock Setup
to
Clock Data Recovery
to get the error detector's
clock from the incoming data stream.
8
Enable the pattern generator outputs by pressing the
0V (Disable)
button.
6
Advanced Analysis
250
Agilent J-BERT N4903B High-Performance Serial BERT
Summary of Contents for J-BERT N4903B
Page 1: ...S Agilent J BERT N4903B High Performance Serial BERT User Guide s Agilent Technologies ...
Page 10: ...10 Agilent J BERT N4903B High Performance Serial BERT ...
Page 36: ...1 Planning the Test 36 Agilent J BERT N4903B High Performance Serial BERT ...
Page 60: ...2 Setting up External Instrument s 60 Agilent J BERT N4903B High Performance Serial BERT ...
Page 120: ...3 Setting up Patterns 120 Agilent J BERT N4903B High Performance Serial BERT ...
Page 360: ...6 Advanced Analysis 360 Agilent J BERT N4903B High Performance Serial BERT ...
Page 468: ...8 Jitter Tolerance Tests 468 Agilent J BERT N4903B High Performance Serial BERT ...
Page 524: ...9 Solving Problems 524 Agilent J BERT N4903B High Performance Serial BERT ...
Page 566: ...10 Customizing the Instrument 566 Agilent J BERT N4903B High Performance Serial BERT ...