Waveform
A graph indicates whether the chosen setting of frequency and
amplitude is tolerable.
N O T E
The lighter area marks the overprogramming area. These values can be
programmed, but they are not calibrated and specified. It is advised to verify error-
free instrument operation before doing the measurement.
The maximum peak-to-peak
Amplitude
is limited by the free capacity of
the chosen delay line.
The maximum
Frequency
of the periodic jitter source depends on the
chosen waveform. Refer to the technical specifications.
You can switch between sine, rectangular, or triangular jitter source
waveforms.
External Jitter Source Parameters
When you enable the
External
checkbox, the
Jitter Setup
function reserves
the remaining free capacity of the 200
+
ps delay line for the signal
applied to the Delay Ctrl Input.
The parameter window informs you about
•
Amplitude (max)
•
Voltage (max)
•
Slope
Jitter Tolerance Tests
7
Agilent J-BERT N4903 High-Performance Serial BERT
333
Amplitude
Frequency
Waveform
Summary of Contents for J-BERT N4903
Page 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Page 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Page 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Page 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Page 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Page 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Page 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...