G4208A User’s Guide
106
2
Working with the Instant Pilot
Simultaneous Execution with Software
With 3
rd
Party Control Software
There may be problems when connecting the Agilent Instant Pilot G4208A to an
Agilent 1100/1200//1260/1290 instrument controlled by third party software, if
this software generates an error when detecting an unknown module in the
configuration. Therefore the Agilent Instant Pilot G4208A can be made invisible
to other controllers.
To change the setting open the
Configure> Controller
screen. Scroll to
3rd Party
Software
and change to
ON
.
Figure 78
Configure Controller - 3rd Party Software Compatibility
Backward compatible OFF
visible to other controllers (default)
Backward compatible ON
invisible to other controllers
NOTE
Whenever this setting is changed, the Instant Pilot needs to be rebooted to make
the change active.
If the Instant Pilot is in
Backward compatible mode ON
on page 106, it will not be seen by the LAN/RS-232 Firmware
Update Tool. For updating firmware reconfigure the Instant Pilot to
Backward
compatible mode OFF
.