Agilent InfinityLab ELSD User Manual
23
2
Site Requirements and Specifications
Performance Specifications
Performance Specifications
Performance Specifications (G4260B)
Table 4
Performance Specifications G4260B
Type
Specification
Comment
Light Source
LED 480 nm
(Class 1 LED product)
Detector
PMT with digital signal
processing
Nebulizer
OFF, 25 – 90 °C
Evaporator
Non-cooled
OFF, 25 – 120 °C
Cooled
Not Applicable
Gas Flow Range
0.9 – 3.25 SLM
(controlled gas shut-off)
Short Term Noise
<
0.2 mV
See conditions in
Drift
<
1 mV/h
See conditions in
Operating Pressure
4.1 – 6.9 bar (60 – 100 psi)
Eluent Flow range
0.2 – 5.0 mL/min
Digital Output
10, 40 or 80 Hz ( 24 bit)
Analogue Output
0 – 1.25 V FSD
Communication
Ethernet
Serial (RS232)
Remote Start Input
Pump Stop: 1 Contact closure
Remote operation
Remote Start Input