Specification Conditions
ASTM: “Evaporative Light Scattering Detectors Used in Liquid Chromatography”.
ASTM drift tests require a temperature change below 2 °C/h (3.6 °F/h) over one
hour period. Our published drift specification is based on these conditions. Larger
ambient temperature changes will result in larger drift. Better drift performance
depends on better control of the temperature fluctuations. To realize the highest
performance, minimize the frequency and the amplitude of the temperature
changes to below 1 °C/h (1.8 °F/h). Turbulences around one minute or less can
be ignored. ASTM measurements require that the detector should be turned on
enough time before start of testing.
PC control
ELSD driver for OpenLab
ChemStation edition
ELSD driver for OpenLab
EZChrom edition
LC and CE Drivers Rev. A.02.11
or above
Safety and maintenance
Gas shut-off Valve, Leak
Detection, Laser Interlock
Table 4
Performance Specifications G4260B
Type
Specification
Comment
NOTE
In order to control OpenLab CDS edition, the ELSD driver is required.
Table 5
Reference conditions
Gas flow
1.6 SLM
Neb temperature
70 °C
Evaporator temperature
70 °C
PMT Gain
1
Data Rate
40 Hz
Smoothing
G4260B: (30)
Light Source intensity
100 %