6000 Series Oscilloscope User’s Guide
2
Front-Panel Controls
For a 4-channel oscilloscope, connect BNC tees to the
equal-length cables as shown below. Then connect a
BNC(f)-to-BNC(f) (barrel connector) to the tee as shown
below.
Figure 11
User Calibration cable for 4-channel oscilloscope
1
Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
2
Press the
Utility
key, then press the
Service
softkey.
3
Begin the Self Cal by pressing the
Start
User Cal
softkey.
4
When the User Cal is completed, set the rear-panel
CALIBRATION switch to PROTECTED.
User Cal Status
Pressing
Utility
&
Service
&
User Cal Status
displays the summary
results of the previous User Cal, and the status of probe
calibrations for probes that can be calibrated. Note that passive
Longer cable
to TRIG OUT
To Channel 3
To Channel 1
To Channel 4
To Channel 2