:TRIGger:FLEXray:FRAMe:TYPE
Command Syntax
:TRIGger:FLEXray:FRAMe:TYPE <frame_type>
<frame_type> ::= {NORMal | STARtup | NULL | SYNC | NSTArtup | NNULl |
NSYNc | ALL}
The :TRIGger:FLEXray:FRAMe:TYPE command sets the FlexRay frame type
to trigger on. The frame type setting is only valid when the FlexRay trigger
mode is set to FRAME.
•
NORMal — will trigger on only normal (NSTArtup & NNULl & NSYNc)
frames.
•
STARtup — will trigger on only startup frames.
•
NULL — will trigger on only null frames.
•
SYNC — will trigger on only sync frames.
•
NSTArtup — will trigger on frames other than startup frames.
•
NNULl — will trigger on frames other than null frames.
•
NSYNc — will trigger on frames other than sync frames.
•
ALL — will trigger on all FlexRay frame types.
Query Syntax
:TRIGger:FLEXray:FRAMe:TYPE?
The :TRIGger:FLEXray:FRAMe:TYPE? query returns the current frame type
setting for the FlexRay frame trigger setup.
Return Format
<frame_type><NL>
<frame_type> ::= {NORM | STAR | NULL | SYNC | NSTA | NNUL |
NSYN | ALL}
See Also
•
"Introduction to :TRIGger Commands"
•
•
N O T E
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...