500
Agilent InfiniiVision 7000A Series Oscilloscopes Programmer's Guide
5
Commands by Subsystem
:TRIGger[:EDGE]:COUPling
Command Syntax
:TRIGger[:EDGE]:COUPling <coupling>
<coupling> ::= {AC | DC | LFReject}
The :TRIGger[:EDGE]:COUPling command sets the input coupling for the
selected trigger sources. The coupling can be set to AC, DC, or LFReject.
•
AC coupling places a high- pass filter (10 Hz for analog channels, and
3.5 Hz for all External trigger inputs) in the trigger path, removing dc
offset voltage from the trigger waveform. Use AC coupling to get a
stable edge trigger when your waveform has a large dc offset.
•
LFReject coupling places a 50 KHz high- pass filter in the trigger path.
•
DC coupling allows dc and ac signals into the trigger path.
Query Syntax
:TRIGger[:EDGE]:COUPling?
The :TRIGger[:EDGE]:COUPling? query returns the current coupling
selection.
Return Format
<coupling><NL>
<coupling> ::= {AC | DC | LFR}
See Also
•
"Introduction to :TRIGger Commands"
•
•
N O T E
The :TRIGger[:EDGE]:COUPling and the :TRIGger[:EDGE]:REJect selections are coupled.
Changing the setting of the :TRIGger[:EDGE]:REJect can change the COUPling setting.
Summary of Contents for InfiniiVision 7000A Series
Page 1: ...Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide...
Page 34: ...34 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 1 What s New...
Page 44: ...44 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 58: ...58 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 750: ...750 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide 8 Error Messages...
Page 922: ...922 Agilent InfiniiVision 7000A Series Oscilloscopes Programmer s Guide Index...