:MEASure Commands
19
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
395
:MEASure:COUNter
Command Syntax
:MEASure:COUNter [<source>]
<source> ::= {<digital channels> | CHANnel<n> | EXTernal}
<digital channels> ::= DIGital<d> for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
The :MEASure:COUNter command installs a screen measurement and
starts a counter measurement. If the optional source parameter is
specified, the current source is modified. Any channel except Math may be
selected for the source.
The counter measurement counts trigger level crossings at the selected
trigger slope and displays the results in Hz. The gate time for the
measurement is automatically adjusted to be 100 ms or twice the current
time window, whichever is longer, up to 1 second. The counter
measurement can measure frequencies up to 125 MHz. The minimum
frequency supported is 1/(2 X gate time).
The Y cursor shows the the edge threshold level used in the measurement.
Only one counter measurement may be displayed at a time.
Query Syntax
:MEASure:COUNter? [<source>]
The :MEASure:COUNter? query measures and outputs the counter
frequency of the specified source.
Return Format
<source><NL>
<source> ::= count in Hertz in NR3 format
See Also
•
"Introduction to :MEASure Commands"
•
•
•
N O T E
This command is not available if the source is MATH.
N O T E
The :MEASure:COUNter? query times out if the counter measurement is installed on the
front panel. Use :MEASure:CLEar to remove the front-panel measurement before executing
the :MEASure:COUNter? query.
Summary of Contents for InfiniiVision 3000 X-Series
Page 1: ...s1 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 30: ...30 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide...
Page 54: ...54 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 68: ...68 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 218: ...218 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 6 Root Commands...
Page 242: ...242 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 280: ...280 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 11 DEMO Commands...
Page 288: ...288 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 12 DIGital d Commands...
Page 340: ...340 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 15 FUNCtion Commands...
Page 358: ...358 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 16 HARDcopy Commands...
Page 378: ...378 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 18 MARKer Commands...
Page 446: ...446 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 19 MEASure Commands...
Page 564: ...564 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 23 POWer Commands...
Page 842: ...842 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 29 TIMebase Commands...
Page 1054: ...1054 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 35 Error Messages...
Page 1076: ...1076 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide 36 Status Reporting...
Page 1232: ...1232 Agilent InfiniiVision 3000 X Series Oscilloscopes Programmer s Guide Index...