85420E/85460A High Frequency Filter Assembly Replacement Kit 3
High Frequency Filter Assembly Replacement Kit
Introduction
This kit contains a replacement HF Filter Assembly (A9).
Installation Kit Parts List
High Frequency Filter Assembly Replacement Kit Contents
Tools Required:
T-10 TORX Driver, T-15 TORX Driver, 5/16 Open-end Wrench
Calibration and Verification
Refer to the 85460A and 85420E Service Guide for the appropriate adjustment routines and
performance verification tests that must be run after replacing the HF Filter Assembly. (Refer to
Chapter 2 “Making Adjustments” Table 2-2, A9 High Frequency Filters Board Assembly.)
Products Affected: . . . . . . . . . . . . . . . . . . . . . . . . .
HP/Agilent 85420E
HP/Agilent 85460A
Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . . .
All
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
none
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . .
(X) Agilent Service Center
(X) Personnel Qualified by Agilent Technologies
Estimated Installation Time: . . . . . . . . . . . . . . . .
1.0 Hours
Estimated Verification Time: . . . . . . . . . . . . . .
5.0 Hours
Item
Quantity
Description
Part Number
1
1
HF Filter Assembly
85460-60030
2
1
Installation Note
85460-90030