![Agilent Technologies G7117C User Manual Download Page 122](http://html1.mh-extra.com/html/agilent-technologies/g7117c/g7117c_user-manual_2867751122.webp)
Agilent InfinityLab LC Series Diode Array Detectors User Manual
122
8
Test Functions and Calibration
Introduction
Introduction
All tests are described based on the Agilent Lab Advisor Software B.02.08. Other
user interfaces may not provide any test or just a few.
For details on the use of the interface refer to the interface documentation.
The Lab Advisor shows the available test under Service & Diagnostics.
Table 21
Interfaces and available test functions
Interface
Comment
Available Function
Agilent Lab Advisor
For functions, refer to Function
Overview Lab Advisor
•
G7117A(
Available functions depend on
Product Level (Basic –
Advanced – FSE)
Agilent ChemStation
No tests available
Adding of temperature/lamp
signals to chromatographic sig
-
nals possible
•
Temperature main board
•
Temperature optical unit
•
Lamp anode voltage
Table 22
Function Overview Lab Advisor Basic/Advanced (G7117A/G7117B/G7117C)
Functions
Product Level
Tests
- ASTM Drift and Noise Test
Basic
Advanced
- Cell Test
Basic
Advanced
- D/A Converter Test
Basic
Advanced
- Dark Current Test
Basic
Advanced
- Intensity Test
Basic
Advanced
- Quick Noise Test
Basic
Advanced
- Self Test
Basic
Advanced
- Slit Test (G7117B only)
Basic
Advanced
- Wavelength Verification Test
Basic
Advanced
Calibrations
- Wavelength Calibration
Basic