90
5977 Series MSD Operation Manual
4
Operating in Chemical Ionization (CI) Mode
General Guidelines
•
Always use the highest purity methane (and other reagent gases, if
applicable). Methane must be at least 99.9995% pure.
•
Always verify the MSD is performing well in EI mode before switching to CI.
See
“To Verify System Performance”
•
Make sure the CI ion source and GC/MSD interface tip seal are installed.
•
Make sure the reagent gas plumbing has no air leaks. This is determined in
PCI mode, checking for
m/z
32 after the methane pretune.
•
Make sure the reagent gas inlet line is equipped with gas purifiers (not
applicable for ammonia).
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......