5977 Series MSD Operation Manual
2
Installing GC Columns
Columns
Many types of GC columns can be used with the MSD but there are some
restrictions.
During tuning or data acquisition, the rate of column flow into the MSD should
not exceed the maximum recommended flow. Therefore, there are limits to
column length and flow. Exceeding recommended flow results in degradation
of mass spectral and sensitivity performance.
Remember that column flows vary greatly with oven temperature. See
Calibrate Column Flow Linear Velocity”
66 for instructions on how to
measure actual flow in your column. Use the Flow Calculation software and
to determine whether a given column will give acceptable flow with
realistic head pressure.
Table 5
Gas flows
Model
Feature
G7035A
G7036A
G7037A
G7038A
G7039A
G7040A
High vacuum pump
Diffusion
Performance
turbo
Diffusion
Performance
turbo
Performance
turbo
Performance
turbo
Optimal He column flow
mL/min
1
1 to 2
1
1 to 2
1 to 2
1 to 2
Maximum recommended gas
flow mL/min
*
1.5
4
1.5
4
4
4
Maximum gas flow, mL/min
†
2
6.5
2
6.5
6.5
6.5
Max column id
0.25
mm
(30 m)
0.53 mm
(30 m)
0.25
mm
(30 m)
0.53 mm
(30 m)
0.53 mm
(30 m)
0.53 mm
(30 m)
CI capability
No
No
No
No
No
Yes
GC compatibility
7820
7820
7890 Series
7890 Series
7890 Series
7890 Series
*
Total gas flow into the MSD: column flow plus reagent gas flow (if applicable).
†
Expect degradation of spectral performance and sensitivity.
Summary of Contents for G7035A
Page 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Page 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Page 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Page 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Page 171: ......