2
Installation
Assembling the Foreline Hose
Agilent G6697A Oil Free IDP-10 Dry Scroll Pump Installation Guide
17
3
Connect the other end of the foreline hose directly to the GC/MS foreline
gauge adapter as shown in
a
Place the remaining Hose Clamp (1400-3241) on the hose
BEFORE
attaching the hose to the gauge adapter as shown in
.
b
Spray the end of the GC/MS foreline gauge adapter with isopropyl
alcohol (IPA) solvent.
c
Slide the end of the foreline hose over the end of the foreline gauge
adapter.
d
Secure the hose on the foreline gauge adapter by positioning the hose
clamp over the barb portion of the foreline gauge adapter then using a
hand tool, tighten hand tight.
Figure 6.
Attach the foreline hose to the GC/MS foreline gauge adapter, and
secure with Hose Clamp