G6012A Quiet Cover DS
User Information
Agilent Technologies
4
Troubleshooting
Table 4 describes solutions to problems that may occur with
Quiet Cover DS. If your problem is not listed or the solution
listed does not work, consult your pump documentation or
contact your Agilent service representative.
WA R N I N G
Hot surfaces can cause burns. Observe the cautions and warnings
covered elsewhere in this manual.
Table 4
Problems and solutions
Problem
Solution
Pump does not fit in Quiet Cover DS.
•
Check that your pump model is supported. See “Introduction” on
page 5.
•
Check for proper pump orientation and alignment on the vibration
foam. See “Installing the Rough Pump Inside Quiet Cover DS” on
page 17.
Call your Agilent service representative if the pump model, orientation,
and configuration are correct.
Temperature alarm does not turn off.
•
Check that ambient room temperature is below 35 °C (95 °F).
•
Remove the Quiet Cover DS access panels and lid.
•
Ensure there are no air restrictions and that adequate cool air is
available to the Quiet Cover DS.
•
Check that both fans are turning.
Call your Agilent service representative if the lab temperature is within
specifications.
Quiet Cover DS does not turn on.
Check that the power supply is plugged securely into both Quiet Cover
DS and the wall outlet. If it is, call your Agilent service representative.
Do not continue to use your rough pump in the Quiet Cover DS.
Quiet Cover DS poorly muffles the sound of the
pump.
•
Ensure that the lid is fastened tightly to the base and that the pump
access panels are securely in place.
•
Check for proper pump alignment on the vibration foam.
The rough pump oil does not drain completely.
Be sure the kickstand is down, with the bar sitting flat on the floor. If so,
refer to the rough pump documentation for more information.