182
1260 SFC Binary Pump
12
LAN Configuration
Manual Configuration
Manual Configuration
Manual configuration only alters the set of parameters stored in the
non-volatile memory of the module. It never affects the currently active
parameters. Therefore, manual configuration can be done at any time. A
power cycle is mandatory to make the stored parameters become the active
parameters, given that the initialization mode selection switches are allowing
it.
Figure 24
Manual Configuration (Principle)
With Telnet
Whenever a TCP/IP connection to the module is possible (TCP/IP parameters
set by any method), the parameters may be altered by opening a Telnet
session.
1
Open the system (DOS) prompt window by clicking on Windows
START
button and select
“Run...”
. Type “cmd” and press OK.