Agilent G4560A Upgrade Kit
Installation Guide
9
1
Remove the pneumatics assembly from the headspace sampler.
a
Completely loosen the two T- 20 Torx captive thumbscrews that are
inset at the front of the pneumatics assembly (
).
b
Completely loosen the T- 20 Torx captive thumbscrew in the back
area of the pneumatics assembly (
c
Lift the pneumatics assembly straight up, rotate the unit about 90
degrees counterclockwise, and lay it across the back of the headspace
sampler.
d
If desired, disconnect the pneumatics assembly cables at the PCM
and heater locations. The cable harness should remain connected to
Figure 7
Front T-20 Torx thumbscrews
Figure 8
Back T-20 Torx thumbscrew