24
Installation Guide
Agilent G4560A Upgrade Kit
Install the 111 vial model covers
1
Install the right panel assembly.
a
Lower the bottom of the panel into the headspace sampler and slide
the panel assembly into position.
b
Install the two T- 20 Torx screws on the right panel assembly.
2
Install the 111 vial front cover.
a
Slide the front cover down and to the right until the cover is in
position on the headspace sampler.
b
Install the two T- 20 Torx screws on the left side to secure the cover
to the headspace sampler.
3
Install the pneumatics front cover.
a
Lower and slide the pneumatics front cover into place.
b
Install the T- 20 Torx screw shown in
Restore the headspace and GC to operating conditions
1
Move the headspace back to its position next to the GC.
2
Reinstall the transfer line to the GC inlet.
3
Restore all gas flows and plug in the headspace power cord.