9
Install the Capillaries
Install the Valve Head and Connect Capillaries
1
Install the in and out connectors.
•
from sampler to the valve ( Capillary SST 0.12x500mm M4- SL PS- PS
(5500- 1202))
•
from valve to the detector ( Capillary SST 0.12x280mm M4- SL PS- PS
(5500- 1203))
Dji
>c
The
In
port is hydraulically connected to the column inlet ports 1- 8 on the
inner ring while the
Out
port connects to the column outlet ports 1`- 8` on the
outer ring.
2
Install the column inlet and outlet connections.
ports 1-8
for connections from valve to the heat exchanger ( Capillary SST 0.12x130mm M4-SL PS-PS
(5500-1199)) or waste line (Waste tube (G1375-87326))
ports 1`-8`
for connections from column outlet to valve ( Capillary SST 0.12x130mm M4-SL PS-NS LS
(5500-1200)), use fingertight PEEK fittings for connecting the column outlet
;gdbXdajbc
Id]ZVi
ZmX]Vc\Zg