To adjust or replace the nebulizer needle
6
Nebulizer Adjustment Fixture User Guide
To adjust or replace the nebulizer needle
When needed
•
When the needle is plugged. Common symptoms of a plugged needle are:
•
increased LC back pressure
•
off-axis spraying
•
dripping from the nebulizer
•
reference nebulizer needle is not spraying.
•
When the needle appears to be corroded or damaged. See
Figure 3
Corroded needle in nebulizer
Tools needed
•
•
clean gloves
•
pliers or wrench, depending on nebulizer holder
•
long-nose pliers (p/n 7810-0004)
•
3-mm open-end wrench (p/n
8710-2699)
•
wrench or driver, depending on nebulizer style:
•
¼-inch x 5/16-inch open-end wrench (p/n 8710-0510)
•
T6 Torx driver (p/n 8710-2548)
•
1.5-mm ball driver (p/n 8710-1570)