Setting up a trial sequence on the 6890 GC for BCR adjustment
1.
Press the [SEQ] key.
2.
Position the < cursor at PRIORITY METH # and enter
0
(zero).
3.
Scroll to TYPE and press either the [FRONT INJ] or [BACK INJ] key,
depending on where the injector is mounted
4.
Scroll to #INJECTIONS/VIAL and enter
1
.
5.
Scroll to SAMPLES and confirm that it reads
1-1
. If not, fix it.
6.
Scroll to METHOD and enter
0
.
7.
Press [SEQ CONTROL]. Do not start the trial sequence until the
following checks and adjustments have been made.
BCR position adjustments for optimal performance
Figure 8 summarizes the adjustments that you will make.
Figure 8.
Adjustment summary
GC
Support bracket
Tray
1. Adjust finger positions to receive vials
2. Adjust bracket mounting screws to level reader
3. Adjust height screws to provide finger clearance
Finger clearance space
Bar code reader