User Information
Advanced Functions
Stored Method
Description
Use the Stored Method screen to load one of five preset
methods. See
for the parameters stored in each method.
READY
Checkout
OQ/PV-GC
MeOH in H2O
↓
STORED METH.
Default
Table 6
Stored Methods
Parameter
Default
Checkout
OQ/PV-GC
MeOH in H
2
0
OQ/PV-MSD
Heated zones
(°C)
Oven
50
100
100
60
100
Loop
60
110
110
110
150
Tr. Line
70
115
115
120
180
Equilibration
time (min)
GC cycle time
25.0
5.5
2.0
2.0
6.0
Vial eq. time
15.0
7.0
7.0
1.5
5.0
Pressuriz. time
0.20
0.08
0.08
0.20
0.1
Loop fill time
0.20
0.5
0.5
0.20
0.5
Loop eq. time
0.05
0.05
0.05
0.05
0.1
Inject time
1.0
1.0
0.2
0.25
1.0
Vial parameters
First vial
1
1
1
1
1
Last vial
10
1
8
1
9
Shake
off
off
off
off
off
Vial size
20 mL
20 mL
20 mL
20 mL
20 mL
Advanced
functions
Multiple
extraction
off
off
off
on
off
Injections per vial 1
1
1
20
1