Introduction to Headspace and to the G1888 Network Headspace Sampler
User Information
37
Vial pressurization
At the end of the vial equilibration time agitation stops, the vial
moves in-line with the sampling probe, and the mechanical rod
raises the vial onto the sampling probe. The pressurization
valve closes. The sampling probe pierces the septum of the vial.
The pressurization valve opens, allowing gas to enter the vial for
the pressurization time. The valve then closes again. See
Figure 8
Vial pressurization. S1 is closed (vial raised on needle).
Carrier gas flow
Mass
controller
Pressure
transducer
Vent
Sample
loop
To GC
Auxiliary gas flow
for vial pressurization
Pressure
regulator
Pressure
transducer
Flow restrictor
S1
S2
Needle
Headspace vial
Open
Closed
Carrier gas
Auxiliary gas
1 2
3
4
5
6
flow
Toggle
switch
Toggle
switch