The objective of this note is to give some guidance on troubleshooting the
Agilent 1100 series well plate sampler.
Additional information will be added when available. The latest version of this
document can be downloaded from the LSBU homepage.
(http://lsbu.marketing.agilent.com/start/start.asp)
Products - Injection system - G1367A - Technical note
Gather information about the problem
•
Instrument serial number?
•
Instrument firmware revision and user interface revision?
•
When did the problem start?
•
What was done/changed prior to the start of the problem?
•
What errors or error codes are shown in the Chemstation GUI, and in the
current logbook files? (check especially for errors around the start date of
the problem)
•
What errors or error codes are shown in the WPS error logbook?
(Chemstation\HPCHEM\instrument\x\temp\lals.txt)
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......