Troubleshooting and Test Functions
Step Commands
Each movement of the sampling sequence can be done under manual control.
This is useful during troubleshooting where close observation of each of the
sampling steps is required to confirm a specific failure mode or verify
successful completion of a repair.
Each step command actually consists of a series of individual commands
which move the sampler components to predefined positions enabling the
specific step to be done.
In the ChemStation the step commands can be selected from the “Test
Selection Box” in the Diagnosis display. In the Control Module the step
commands can be accessed from the pull-down menu in the well-plate sampler
“Test”.
Table 16
Step Commands
Step
Action
Comments
Bypass
Switches injection valve to the bypass
position.
Plunger Home
Moves the plunger to the home
position.
Needle Up
Lifts the needle arm to the upper
position.
Command also switches the
valve to bypass if it is not
already in that position.
Needle into Vial
Lowers the needle into the vial.
Draw
Metering device draws the defined
injection volume.
Command lifts the needle, and
lowers the needle into vial.
Command can be done more
than once (maximum draw
volume of 100µl cannot be
exceeded). Use “Plunger
Home” to reset the metering
device.
Needle Up
Lifts the needle out of the vial.
Command also switches the
valve to bypass if it is not
already in that position.
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......