Peristaltic Pump Error
The peristaltic pump motor in the well-plate sampler has failed.
The current on the motor is used by the MTP board to monitor the speed of the
peristaltic pump motor. If the current falls below a certain value, the error
message is generated.
Probable Causes
•
Defective motor.
•
Defective SUD board.
•
Defective MTP main board.
Suggested actions
✔
Exchange the peristaltic pump motor.
✔
Exchange the SUD board.
✔
Exchange the MTP main board.