62
1100 Series WS MWS Reference Manual
3
Troubleshooting and Test Functions
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The sampler continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
•
Leak detected in another Agilent 1100 module with a CAN connection to
the system.
•
Error detected in an external instrument with a remote connection to the
system.
•
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
✔
Determine which Agilent 1100 module has the leak. Fix the leak before
restarting the well-plate sampler.
✔
Check external instruments for an error condition.
✔
Check the degasser for an error condition. Refer to the
Reference Manual
for the Agilent 1100 Series degasser.