Theory of Operation
7
1100 Series WS MWS Reference Manual
235
Communication Settings for RS-232C Communication
The communication protocol used in this instrument supports only hardware
handshake (CTS/RTS).
Switches 1 in down and 2 in up position define that the RS-232C parameters
will be changed. Once the change has been completed, the instrument must be
powered up again in order to store the values in the non-volatile memory.
Use the following tables for selecting the setting which you want to use for
RS-232C communication. The number 0 means that the switch is down and 1
means that the switch is up.
Table 71
Communication Settings for RS-232C Communication
Mode Select
1
2
3
4
5
6
7
8
RS-232C
0
1
Baud rate
Data
Bits
Parity
Table 72
Baud Rate Settings
Switches
Baud Rate
Switches
Baud Rate
3
4
5
3
4
5
0
0
0
9600
1
0
0
9600
0
0
1
1200
1
0
1
14400
0
1
0
2400
1
1
0
19200
0
1
1
4800
1
1
1
38400
Table 73
Data Bit Settings
Switch 6
Data Word Size
0
7 Bit Communication
1
8 Bit Communication
Summary of Contents for G1367A
Page 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Page 4: ...4 1100 Series WS MWS Reference Manual...
Page 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Page 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Page 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Page 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Page 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Page 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Page 284: ...284 1100 Series WS MWS Reference Manual Index...
Page 285: ......