Agilent 1260 Infinity VWD User Manual
97
Error Information
7
Detector Error Messages
Wavelength calibration setting failed
Error ID: 7310
The intensity maximum was not found during wavelength calibration.
Calibration 0 Failed:
Zero-order calibration failed.
Calibration 1 Failed:
656 nm calibration failed.
Probable cause
Suggested actions
1
Lamp is OFF.
Switch on the lamp.
2
Incorrect flow cell installation.
Ensure the flow cell is installed correctly.
3
Flow cell contamination or air bubbles.
Clean/replace flow cell windows or remove air
bubbles.
4
Intensity too low.
Replace lamp.
5
Current step value too far from maximum.
•
Repeat the calibration.
•
Please contact your Agilent service
representative.
6
Misaligned/defective grating assembly.
Please contact your Agilent service
representative.
7
Defective main board.
Please contact your Agilent service
representative.