Agilent 1260 Infinity VWD User Manual
169
Hardware Information
12
Optional Interface Boards
Optional Interface Boards
BCD / External Contact Board
The Agilent 1200 Infinity Series modules have one optional board slot that
allows to add an interface board to the modules. Some modules do not
have this interface slot. Refer to
on page 175 for details.
Optional Interface Boards
The BCD board provides a BCD output for the bottle number of the
Agilent 1200 Series autosampler and four external contacts. The external
contact closure contacts are relay contacts. The maximum settings are:
30 V (AC/DC); 250 mA (fused).
p/n
Description
G1351-68701
Interface board (BCD) with external contacts and BCD outputs
2110-0004
Fuse for BCD board, 250 mA
A^cZYg^kZg
&'
)m
'*%b6
G;>[^aiZg
G;>[^aiZg
XdccZXidg
:miZgcVa
XdciVXih
7dVgY
^YZci^[^XVi^dc
EgdXZhhdg
^ciZg[VXZ
789
XdccZXidg
:miZgcVa
XdciVXi
789
gZ\^hiZg