Agilent 1260 Infinity VWD User Manual
7
Error Information
Detector Error Messages
Cutoff filter doesn't decrease the light intensity at 250 nm
Error ID: 7813
The automatic filter check after lamp ignition has failed. When the lamp is
switched on, the detector moves the cutoff filter into the light path. If the
filter is functioning correctly, a decrease in lamp intensity is seen. If the
expected intensity decrease is not detected, the error message is
generated.
ADC Hardware Error
Error ID: 7830, 7831
A/D- Converter hardware is defective.
Probable cause
Suggested actions
1
Motor is not connected.
Please contact your Agilent service
representative.
2
Defective motor.
Please contact your Agilent service
representative.
3
Defective/missing grating or filter.
Please contact your Agilent service
representative.
4
Cable/connector defective.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
A/D-Converter hardware is defective.
Please contact your Agilent service
representative.