Repairing the Autosampler
Power Supply
Power Supply
C A U T I O N
The ASM board is sensitive to electrostatic discharge. Always use the ESD
strap (see “Using the ESD Strap” on page 105) when handling electronic
boards.
Frequency
When defective
Tools required
Pozidrive No. 1 screwdriver
Parts required
Power supply 0950-2528.
Before beginning this procedure:
❏
Remove the top cover (page 128).
❏
Remove the transport assembly (page 132).
❏
Remove the sampling unit (page 134).
❏
Remove the ASM board (page 148).
❏
Remove the fan (page 146).
1
Remove the bottom foam.
2
Remove the power switch light pipe from the
coupler.
3
Remove the two screws from the rear panel.
CellFrame
CellFrame
Light
pipe
Coupler
CellFrame
Summary of Contents for G1313A
Page 1: ...s1 Agilent 1100 Series Standard Micro and Preparative Autosamplers Reference Manual ...
Page 3: ...Reference Manual Agilent 1100 Series Standard Micro and Preparative Autosamplers ...
Page 12: ...12 Contents ...
Page 39: ...2 2 Optimizing Performance How to optimize the autosamplers to achieve best results ...
Page 93: ...93 Troubleshooting and Test Functions Step Commands ...
Page 216: ...216 Introduction to the Autosampler Electrical Connections ...
Page 244: ...244 Theory of Operation The Main Power Supply Assembly ...
Page 265: ...265 Introduction to the Control Module Diagnostics and Tests ...
Page 268: ...268 Introduction to the Control Module Diagnostics and Tests ...
Page 274: ...274 Specifications Performance Specifications ...
Page 275: ...10 Legal Safety and Warranty Information ...
Page 286: ...286 Agilent Technologies on Internet ...
Page 290: ...290 Index ...
Page 291: ......