Repairing the Autosampler
Interface Board
Interface Board
C A U T I O N
The interface board is sensitive to electrostatic discharge. Always use the ESD
strap when handling electronic boards.
1
Switch off the autosampler at the main power switch.
2
Disconnect cables from the interface board connectors.
3
Loosen the screws. Slide out the interface board from the autosampler.
4
Install the interface board. Secure the screws.
5
Reconnect the cables to the board connectors
Figure 16
Exchanging the Interface Board
Frequency
At installation or when defective.
Tools required
Flat-head screwdriver.
Parts required
Interface board, see “Optional Interface Boards” on page 228.
Screws