20
Reference Manual
Flow Tracker Series 1000/2000
GC mode
) displays the following
parameters:
Linear velocity
The Flow Tracker Series flowmeters use the volumetric flow
rate and the selected capillary diameter (inside diameter) to
calculate the average velocity of the selected gas through the
capillary. This parameter is located in the upper left portion of
the display under the dynamic label “LinVel”. The engineering
unit associated with the linear velocity is centimeters per
second (cm/sec). The calculation is as follows:
Linear Velocity = Volumetric Flow Rate / Capillary Area
Figure 3
GC mode screen/buttons
GAS
MODE
OFF
ON
ZERO
Gas select
Power off
Mode
Mode indicator
Low battery indicator
Reference flow
Linear velocity
Capillary diameter
Split ratio
Primary display
(default is linear
velocity)
Power on
Zero (tare)