Agilent Technologies FieldFox N9923A Quick Reference Manual Download Page 9

Calibration 

Calibration is performed in NA, CAT, and VVM Modes. 

A CalReady calibration is automatically applied to 
measurements when the FieldFox is powered ON and 
when Preset. This calibration sets the reference plane 
at the test port connectors. 

Calibration should be performed when any of the 
following occur: 

 

The temperature changes significantly. 

 

The Frequency Range, Resolution, Min Swp Time, 
or Power Level is changed. 

 

When using a phase stable jumper cable or adapter 
to connect the DUT to the FieldFox. These should 
be high-quality components.  

After making measurement settings (see NA Mode or 
CAT Mode measurements): 

 

Press 

Cal   5

 then follow the prompts. 

 

For highest accuracy with minimum connections 
for your setup, perform the Calibration Type that 
is recommended by the FieldFox. 

 

During a QuickCal, press 

Skip Load

 when NOT 

using a jumper cable or adapter. 

Cal Rdy

 is shown on the screen when 

CalReady

 is 

correcting the measurement.  

Cal ON Q  

is shown on the screen when a 

QuickCal

 is 

correcting the measurement. 

CAL ON U 

is shown on the screen when a Response 

Cal or Mechanical Cal is correcting the measurement. 

When a calibration is being interpolated, an asterisk is 
added to the screen annotation as follows: 

Cal ON Q* 

or  

CAL ON U*

 

Summary of Contents for FieldFox N9923A

Page 1: ...2 The Power Button and LED 2 Battery Usage 3 Measure S Parameters 4 Multi Trace Configurations 6 Measure DTF 7 Measure Cable Loss 1 Port 8 Calibration 9 Power Meter Mode 10 Save and Recall Files 11 Manufacturing Part Number N9923 90002 Print Date February 19 2010 Supersedes None Agilent Technologies Inc 2010 ...

Page 2: ...ive Save files to these devices for quick download to a PC USB Power Sensor For making Power Meter measurements Opt 302 Battery charger or extra battery The Power Button and LED To turn power ON briefly press the power button The LED is green The boot up takes about 60 seconds To switch to Stand by power low battery drain briefly press the power button The LED is amber and blinks slowly To turn po...

Page 3: ...play as much as possible Briefly press the power button to switch to Stand by mode when not being used Press again to restore power All current settings are preserved The FieldFox will shutdown automatically when battery power is very low to prevent it from being completely discharged Use ONLY a FieldFox charger to recharge a battery The battery can be fully charged while in the FieldFox in about ...

Page 4: ...asurement Requires Opt 122 OR Press Trace 6 then Num of Traces then select a Multi trace configuration Learn more on page 6 3 Press Freq Dist then either Start and Stop or Center and Span to enter a frequency range for the measurement 4 Press BW 2 then IFBW to select the IF Bandwidth for the measurement Narrower bandwidths require more time to sweep 5 Press Sweep 3 then Resolution to select the nu...

Page 5: ...Scale Manually enter a scale per division to view specific areas of the trace Ref Level Manually set the value of the reference line Ref Position Manually set the position of the reference line Values must be between 0 TOP line and 10 BOTTOM line Press Measure 1 then Format to change the format for the measurement Press BW 2 then Averaqe then enter a value for the number of trace sweeps to average...

Page 6: ...h as format markers and limit lines are applied to the ACTIVE trace in the same manner as when a single trace is present To select a multi trace configuration Press Trace 6 then Num Traces Then choose a multi trace configuration To activate a trace in order to change a setting Use the arrows to scroll through the traces OR Press Trace 6 then select Trace 1 Trace 2 Trace 3 or Trace 4 Only traces th...

Page 7: ...stance to Fault 5 Press Freq Dist then Stop Distance and enter the length of the DUT 6 If a diplexer or other filter is in the measurement path press Freq Dist and manually type the Min Start and Max Stop frequencies Otherwise press Meas Setup then select Frequency Mode Lowpass 7 Press Cal 5 and follow the Cal prompts 8 Disconnect any components or antenna that should NOT be measured and connect a...

Page 8: ...lues of the measurement 6 Press Sweep 3 then Min Swp Time Increase the sweep time until a stable trace is visible 7 Press Cal 5 then follow the calibration prompts 8 Connect the DUT cable to be tested to the FieldFox and connect a LOAD at the end of the DUT Depending on the DUT the following three steps may improve the amount of ripple that is visible on the screen 9 Press Trace 6 then Data Mem 10...

Page 9: ...ould be high quality components After making measurement settings see NA Mode or CAT Mode measurements Press Cal 5 then follow the prompts For highest accuracy with minimum connections for your setup perform the Calibration Type that is recommended by the FieldFox During a QuickCal press Skip Load when NOT using a jumper cable or adapter Cal Rdy is shown on the screen when CalReady is correcting t...

Page 10: ...ollows Select Offset Press Scale Amptd then Offset On Off Then Offset Val Enter a value by which the power meter display is offset A positive value compensates for a component with loss such as an attenuator Zeroing The Agilent USB Power Sensors perform internal zeroing automatically Internal zeroing does NOT require that the power source be turned OFF Perform external zeroing when measuring power...

Page 11: ...Recalls files to or from a USB Flash drive 2 Then File Type State Saves ALL FieldFox instrument settings and calibration for ALL Modes to an sta file Trace State Saves the current trace all traces in SA mode and instrument settings to an sta file Picture Saves the FieldFox screen to a png file Data S1P Saves CAT and NA Mode trace data to an S1P file Data S2P for NA mode only Data CSV Saves trace d...

Reviews: