Eyesys Mini-B/A Gauge Controller
12
Voltage Output
Pressure signal analog voltage is available between the Signal Out line (pin 9) and Signal
Ground line (pin 10) of J1. The voltage is a 1V/decade output representing the Log of the
pressure, offset to a range between 1 and 9 VDC (Table 7).
0.5 V output indicates pressure below operating range of the Eyesys Mini-BA
Figure 5
Output Voltage as a Function of Pressure
Table 7
Voltage Outputs
Voltages greater than
But less than
Indicate pressures in this range
1
2
1.33x10
-7
Pa, 1x10
-9
Torr, 1.33x10
-9
mbar
2
3
1.33x10
-6
Pa, 1x10
-8
Torr, 1.33x10
-8
mbar
3
4
1.33x10
-5
Pa, 1x10
-7
Torr, 1.33x10
-7
mbar
4
5
1.33x10
-4
Pa, 1x10
-6
Torr, 1.33x10
-6
mbar
5
6
1.33x10
-3
Pa, 1x10
-5
Torr, 1.33x10
-5
mbar
6
7
1.33x10
-2
Pa, 1x10
-4
Torr, 1.33x10
-4
mbar
7
8
1.33x10
-1
Pa, 1x10
-3
Torr, 1.33x10
-3
mbar
8
9
1.33x100 Pa, 1x10
-2
Torr, 1.33x10
-2
mbar
Summary of Contents for Eyesys Mini-B/A
Page 12: ...Eyesys Mini B A Gauge Controller This page intentionally left blank ...
Page 14: ...Eyesys Mini B A Gauge Controller This page intentionally left blank ...
Page 28: ...Eyesys Mini B A Gauge Controller 14 Table 8 Gas Correction Factors ...
Page 29: ...Eyesys Mini B A Gauge Controller 15 Table 7 Gas Correction Factors continued ...
Page 30: ...Eyesys Mini B A Gauge Controller This page intentionally left blank ...
Page 39: ...Returned Materials Report ...
Page 40: ......
Page 41: ......