![Agilent Technologies EXSLRSH110 Instruction Manual Download Page 8](http://html1.mh-extra.com/html/agilent-technologies/exslrsh110/exslrsh110_instruction-manual_2867723008.webp)
SH-110 Exhaust Silencer Kit
1-2
D
R
A
F
T
Installation
NOTE
To ensure proper operation, and to prevent damage, the kit
must be installed and operated only as specified.
To install:
1. Unscrew the ¼" NPT coupler for the pump's exhaust (unless using the small footprint
version).
2. Screw in the ¼" NPT hose barb fitting into the ¼" NPT female exhaust fitting.
Refer to Figure 1-1 for this procedure.
Figure 1-1
: SH-110 Exhaust Silencer Kit: Components
3. Connect the longer piece of Tygon tubing into the ½" hose end of the 1/4" NPT barb and
secure with a hose clamp.
4. Connect the other end of the Tygon tubing to the ½" hose, 3/8" NPT barb belonging to
the Resonator assembly and secure with a hose clamp.
5. Connect the second Tygon tubing to the ½" hose end of the 1/4" NPT barb and secure
with a hose clamp.
6. Attach the smooth end of the NW16 connector to the other end of the Tygon tubing.
7. Connect the NW16 flange to the NW16 connector by the means of a centering ring and
secure connection with a clamp.
Optional: if a smaller footprint is desired (Figure 1-4 on page 1-3), connect the
Silencer-Filter Can NW16 connector directly to the pump's exhaust, which has an NW
16 flange, and secure it with a clamp via a centering ring.
Summary of Contents for EXSLRSH110
Page 2: ...D R A F T Copyright 2008 Agilent Inc SH 110 Exhaust Silencer Kit...
Page 4: ...SH 110 Exhaust Silencer Kit This page intentionally left blank D R A F T...
Page 6: ...SH 110 Exhaust Silencer Kit This page intentionally left blank D R A F T...
Page 14: ...SH 110 Exhaust Silencer Kit This page intentionally left blank...
Page 15: ......