Agilent E8663B Analog Signal Generator User’s Guide
65
5
Troubleshooting
This chapter provides basic troubleshooting information for the Agilent E8663B analog signal
generator. If you do not find a solution here, refer to the
E8663B Analog Signal Generator Service
Guide
.
NOTE
If the signal generator displays an error, always read the error message text by pressing
Utility
>
Error Info
.
•
“RF Output Power Problems” on page 65
•
“No Modulation at the RF Output” on page 69
•
•
“Data Storage Problems” on page 71
•
“Cannot Turn Off Help Mode” on page 72
•
“Signal Generator Locks Up” on page 72
•
•
“Contacting Agilent Sales and Service Offices” on page 75
•
“Returning a Signal Generator to Agilent Technologies” on page 75
RF Output Power Problems
Check the
RF ON/OFF
annunciator on the display. If it reads
RF OFF
, press
RF On/Off
to toggle the RF
output on.
No RF Output Power when Playing a Waveform File
Preset the signal generator, then replay the waveform file.
If a header file is not specified for a waveform, the signal generator uses a default header file with
unspecified settings. If you play a waveform file that has unspecified signal generator settings
(settings not saved to the file header, see
), the signal generator will use the header file
settings from the previously played file. If the previous header file had a marker routed to RF
blanking, the RF output power will be blanked. Preset the signal generator to return the RF blanking
marker function to its default state—off). Refer to the
E8663B Analog Signal Generator Key
Reference
, Marker section, for more information.
NOTE
If the default marker file is used, ensure that the
Pulse/RF Blank
softkey is set to
None
. Markers
may have been set to
Pulse/RF Blank
by a previous file header.