Service Guide E8364-90026
4-17
PNA Series Microwave Network Analyzers
Troubleshooting
E8362B, E8363B, E8364B
Front Panel Troubleshooting
Front Panel Troubleshooting
The front panel assembly consists of the A1 keypad, A2 display, and the A3 front panel
interface. The following tests verify the operation of the front panel assembly when the
analyzer is in the measurement mode. If the instrument fails to power up correctly, or it is
difficult to verify due to a faulty display, refer to
“Power Up Troubleshooting” on page 4-6
.
Refer to the following sections to verify the operation of the noted assemblies.
•
“A1 Front Panel Keypad and RPG Test”
below
•
“A2 Display Test” on page 4-19
•
“A3 Front Panel Interface Board” on page 4-20
— USB
— Speaker
— Probe Power
If all assemblies are working correctly, continue troubleshooting with
A1 Front Panel Keypad and RPG Test
Test the front panel keypad by running the front panel test. To run the front panel test,
perform the following:
On the
System
menu, point to
Service
and click
Front Panel Test.
A
Front Panel Service Test
dialog box will be displayed, as shown in
Figure 4-7 Front Panel Service Test Dialog Box
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...