Service Guide E8364-90026
8-11
PNA Series Microwave Network Analyzers
General Purpose Maintenance Procedures
E8362B, E8363B, E8364B
Error Terms
IMPORTANT
Data are listed here as a convenience only. Detailed instrument specifications
are listed in the embedded help in the network analyzer.
If Error Terms Seem Worse than Expected
To verify that the system still conforms to specifications, perform a system verification.
Refer to
Chapter 3, “Tests and Adjustments”
.
Table 8-1
Error Term Data
a
a. Uncorrected system performance. The values in the table apply over an environmental
temperature of 23 °C ±3 °C, with less than 1 °C deviation from the calibration
temperature.
Parameter
Frequency Range
10 MHz
to
45 MHz
b
b. Typical performance.
45 MHz
to
2 GHz
2 GHz
to
10 GHz
10 GHz
to
20 GHz
20 GHz
to
40 GHz
40 GHz
to
45 GHz
45 GHz
to
50 GHz
Directivity
>22 dB
>24 dB
>22 dB
>16 dB
>16 dB
>15 dB
>13 dB
Source Match
Standard
Option UNL, 014,
or UNL & 014
>17 dB
>12 db
>23 dB
>18 dB
>16 dB
>14 dB
>14 dB
>12 dB
>10 dB
>8 dB
>9 dB
>7 dB
>5.5 dB
>4 dB
Load Match
Standard
Option UNL, 014,
or UNL & 014
>20 dB
>12 dB
>23 dB
>17 dB
>14 dB
>13 dB
>10 dB
>10 dB
>9 dB
>9 dB
>9 dB
>8 dB
>7 dB
>6 dB
Reflection Tracking
±1.5 dB
±1.5 dB
±1.5 dB
±1.5 dB
±1.5 dB
±2.0 dB
±2.0 dB
Transmission
Tracking
c
c. Transmission tracking performance is strongly dependent on cable used. These typical
specifications are set based on the use of Agilent through cable part number 85133-60016.
±2.0 dB
±1.5 dB
±2.0 dB
±2.5 dB
±3.5 dB
±4.0 dB
±4.5 dB
Crosstalk (Isolation)
Standard
Option UNL or 014
Option UNL & 014
Option 080 enabled
d
d. 0 Hz offset.
-85 dB
-85 dB
-85 dB
-85 dB
-100 dB
e
-100 dB
-98 dB
-108 dB
f
e. The value is -85 dB for frequencies below 1 GHz.
f. The value is -95 dB from 45 MHz to 1 GHz.
-110 dB
-109 dB
-108 dB
-115 dB
-110 dB
-109 dB
-107 dB
-120 dB
-108 dB
-106 dB
-104 dB
-115 dB
-105 dB
-103 dB
-100 dB
-112 dB
-100 dB
-98 dB
-95 dB
-105 dB
Summary of Contents for E8362B
Page 11: ...Service Guide E8364 90026 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90026 2 1 2 General Product Information ...
Page 33: ...Service Guide E8364 90026 3 1 3 Tests and Adjustments ...
Page 83: ...Service Guide E8364 90026 4 1 4 Troubleshooting ...
Page 151: ...Service Guide E8364 90026 5 1 5 Theory of Operation ...
Page 185: ...Service Guide E8364 90026 6 1 6 Replaceable Parts ...
Page 269: ...Service Guide E8364 90026 7 1 7 Repair and Replacement Procedures ...
Page 351: ...Service Guide E8364 90026 8 1 8 General Purpose Maintenance Procedures ...