Service Guide E8364-90001
4- 11
PNA Series Microwave Network Analyzers
Troubleshooting
E8362A, E8363A, E8364A
Power Up Troubleshooting
Isolate the remaining three assemblies:
• remove the A15 CPU board
• remove the A4 power supply
• This leaves only the A14 system motherboard installed. If the measurements are still
incorrect, this is the suspected faulty assembly.
Table 4-2
Power Supply Voltages and Resistances
Measurement
Location
Supply
(Vdc)
Approximate
Resistance (
Ω
)
A
+5.0
285
B
-15.0
12.7 k
C
-5.0
7.2 k
D
+9
4.0 k
E
+15
3.0 k
F
+22
2.8 k
G
+22
890
H
+32
2.6 k