3-6
Service Guide E8364-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
About System Verification and Performance Tests
E8362A, E8363A, E8364A
About System Verification and Performance Tests
The performance of the network analyzer is specified in two ways: system specifications,
and instrument specifications. Respectively, the analyzer’s conformance to these
specifications is verified in two ways: system verification, and performance tests.
System Specifications
System specifications specify warranted performance of the measurement system when
making error-corrected measurements. The measurement system includes the analyzer,
test cables, and calibration kit.
The analyzer's system specifications are described in the
Agilent PNA Series Network
Analyzer Data Sheet
and also in the analyzer’s on-line help system.
System specifications are included in section titled “Corrected System Performance.”
System specifications are expressed in two ways:
• residual errors of the measurement system
• graphs of measurement uncertainty versus reflection and transmission coefficients
System specifications are applicable when the measurement system is used to make
error-corrected measurements.
System specifications are verified in one of the following ways:
• Complete the system verification procedure using a certified verification kit, or
• Complete all of the performance tests and certify (or recertify) the calibration kit that
will be used for future measurements. This alternative verifies both the system
specifications and the instrument specifications for the analyzer.
Instrument Specifications
Instrument specifications specify the network analyzer's uncorrected measurement port
characteristics and its output and input behavior.
The analyzer's instrument specifications are described in the
Agilent PNA Series Network
Analyzer Data Sheet
and also in the analyzer’s on-line help system. The sections that
describe instrument specifications are titled:
• “Uncorrected System Performance”
• “Test Port Output”
• “Test Port Input”
These specifications apply when the analyzer is used to make measurements other than
error-corrected measurements. An example would be the measurement of amplifier gain
compression.
Performance tests are used to confirm that the analyzer meets the instrument
specifications.
Summary of Contents for E8362A
Page 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Page 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Page 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Page 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Page 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Page 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Page 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Page 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Page 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Page 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Page 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...