290
Chapter 5
Language Reference
[:SENSe]:FREQuency Subsection
[:SENSe]:FREQuency Subsection
Center Frequency
[:SENSe]:FREQuency:CENTer <freq>
[:SENSe]:FREQuency:CENTer UP|DOWN
[:SENSe]:FREQuency:CENTer?
Set the center frequency.
NOTE
In log sweep mode, the minimum start frequency is 10 Hz.
Factory Preset
and *RST:
600 MHz
Range:
EMC E7401A: –80 MHz
1
to 1.58 GHz
EMC E7402A: –80 MHz
to 3.10 GHz
EMC E7403A: –80 MHz
to 6.78 GHz
EMC E7404A: –80 MHz
to 13.3 GHz
EMC E7405A: –80 MHz
to 27.0 GHz
Default Unit:
Hz
Front Panel
Access:
FREQUENCY/Channel, Center Freq
Center Frequency Step Size Automatic
[:SENSe]:FREQuency:CENTer:STEP:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
Specifies whether the step size is set automatically based on the span.
Factory Preset
and *RST:
On
Front Panel
Access:
FREQUENCY/Channel, CF Step Auto Man
1. 10 Hz minimum in log sweep mode.