![Agilent Technologies E5070B User Manual Download Page 532](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065532.webp)
526
Chapter 15
Measurement Examples
Evaluating Transmission Characteristics of a Front End Module
Step 5.
Perform calibration for channels 2, 3, and 4 according to the same procedure as in Step 2 to
Step 4.
NOTE
Because the test port assignment setting for channels 1 and 2 and that for channels 3 and 4
are the same, you need not change the ECal connection.
9. Connecting DUT
Connect the DUT as shown in Figure 15-37.
Figure 15-37
Connecting 6-port front end module