Measurement Examples
Evaluating a Duplexer
Step 2.
Enter the setup data in the limit table for trace 1 (Figure 15-16).
Figure 15-16
Completed limit table for trace 1
Step 3.
Display the limit table for Trace 2 (S
21
).
Setup Description
Key Operation
Activating Trace 2
Step 4.
Enter the setup data in the limit table for trace 2 (Figure 15-17).
Figure 15-17
Completed limit table for trace 2
11. Execute the Limit Test
The limit test is executed.
Step 1.
Turn on the limit line and limit test for Trace 1.
Setup Description
Key Operation
Activating Trace 1
Limit Line: ON
-
Limit Test
-
Limit Line
(Turn it
ON
)
Limit Test: ON
Limit Test
(Turn it
ON
)