Chapter 15
501
Measurement Examples
Evaluating a Duplexer
15
. Me
asur
eme
n
t Exa
m
ple
s
In the example of Figure 15-15, return loss (response value of Marker 1) is 13.80 dB.
Step 6.
Turn on the marker table display.
Setup Description
Key Operation
Marker table display: ON
-
Marker Table
(Turn it
ON
)
Figure 15-15
Analyzing duplexer parameters
10. Define the Setup for a Limit Table
Follow the steps below to make entries in the limit table.
Step 1.
Display the limit table for Trace 1 (S
13
).
Setup Description
Key Operation
Activating Trace 1
Displaying a limit table
-
Limit Test
-
Edit Limit Line
Moving Marker 1 to the peak in the pass band
Peak
-
Search Left
|
Search Right
(press as
many times as necessary)
*1
*1. If you cannot move the marker to the desired peak, use
Peak Excursion
to change the
peak deviation and then execute the search again. (Example:
Peak Excursion
-
).
Setup Description
Key Operation
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...