480
Chapter 14
Controlling E5091A
Setting the E5091A
Setting control line
The E5070B/E5071B can control the output from the control line (Figure 14-5) of the
E5091A and control the DUT (for example, switching the frequency band of the front end
module). The procedure is shown below. For the specifications of the DUT control line,
refer to
E5091A Users & Service Guide
.
Figure 14-5
Control line
Operational procedure
Step 1.
Press
-
E5091A Setup
to display the E5091A setup menu.
Step 2.
Press
or
to activate the channel for which you want to set the
control line.
Step 3.
After the execution of “Selecting ID for E5091A” on page 476 and “Selecting the E5091A
Model” on page 477, press
Control Lines
to display the settings menu for the DUT control
line.
Step 4.
Use the corresponding softkey to set the control line of the E5091A.
Softkey
Function
Line 1
,
Line 2
,
Line 3
,
Line 4
,
Line 5
,
Line 6
,
Line 7
,
Line 8
Set
High
/
Low
of each line of the control line.
Step 5.
Execute Step. 3 through Step. 4 for all channels that you want to sweep.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...