Power Calibration
4. Cal
ibra
tion
Figure 4-62
Example of creating power sensor calibration factor table
Saving power sensor calibration factor table
You can save the power sensor calibration factor table as a CSV (Comma Separated Value)
format file.
Step 1.
Press
.
Step 2.
Press
Power Calibration
.
Step 3.
Press
Sensor A Settings
or
Sensor B Settings
.
Step 4.
Press
Export to CSV File
to open the Save As dialog box. For information on the Save As
dialog box, see the description in Figure 10-9, “Printers window,” on page 370. At this
time, CSV File (extension*.csv) is selected as the file type.
Step 5.
Enter a file name in the
File Name
box and press the
Save
button to save the power sensor
calibration factor table.
Recalling power sensor calibration factor table
By recalling a power sensor calibration factor table saved in the CSV format according to
“Saving power sensor calibration factor table” on page 213, you can set the power sensor
calibration factor table.
Step 1.
Press
.
Step 2.
Press
Power Calibration
.
Step 3.
Press
Sensor A Settings
or
Sensor B Settings
.
Step 4.
Press
Import from CSV File
to open the Open dialog box. For information on the Open
dialog box, see the description in Figure 10-10, “File menu in Printers window,” on
page 371. At this time, CSV File (extension*.csv) is selected as the file type.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...