Chapter 4
205
Calibration
Specifying Different Standard for Each Frequency
4. Cal
ibra
tion
Notes on how frequency ranges are dealt when using subclasses
By using several subclasses, you can set a different standard for each frequency, but note
the following on frequency ranges. The following table shows possible cases and whether
calibration is possible.
Table 4-7
Possible cases of frequency ranges and whether calibration is possible
Possible cases of frequency ranges
Execution of
calibration
Example 4-1, “When frequency ranges specified with subclasses
overlap,”
Possible
Impossible
Possible
Possible
Example 4-1
When frequency ranges specified with subclasses overlap
For the measurement frequency range (1 GHz - 8 GHz), if you execute calibration with a
standard of subclass 1 (1 GHz-5 GHz) and then with a standard of subclass 2 (4 GHz-8
GHz), the standard last executed is applied for the overlapping portion (4 GHz-5 GHz).
Figure 4-57
When frequency ranges overlap
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...