Chapter 4
145
Calibration
4-port TRL calibration
4. Cal
ibra
tion
Figure 4-35
Connection of standards in 4-port TRL calibration
Operational procedure
NOTE
The shapes of actual standards differ from those shown in Figure 4-35 which are just
symbols for illustration.
Step 1.
Press
or
to select the channel for which you want to execute
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
4-Port TRL Cal
.
Step 5.
Press
Select Ports
.
Step 6.
Thru/Line
to start the measurement of the calibration standard.
Softkey
Function
Port 1-2 Thru
Executes thru/line calibration for test ports 1 and 2.
Port 1-3 Thru
Executes thru/line calibration for test ports 1 and 3.
Port 1-4 Thru
Executes thru/line calibration for test ports 1 and 4.
Port 2-3 Thru
Executes thru/line calibration for test ports 2 and 3.
Port 2-4 Thru
Executes thru/line calibration for test ports 2 and 4.
Port 3-4 Thru
Executes thru/line calibration for test ports 3 and 4.
Summary of Contents for E5070B
Page 6: ......
Page 30: ...24 Contents ...
Page 34: ...28 Chapter1 Precautions Before contacting us ...
Page 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Page 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Page 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Page 634: ...628 AppendixB Troubleshooting Warning Message ...
Page 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Page 740: ...734 AppendixE General Principles of Operation Data Processing ...
Page 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...